Zeta potential measurement
Cordouan WALLIS
The zeta potential describes the interparticle interaction and allows for an assessment of the temporal stability of suspensions.
In the measurement principle used, Laser Doppler Electrophoresis, the particles are set into vibration by applying an alternating electric field. This movement causes a Doppler shift of the scattered light, from which the electrophoretic mobility and the zeta potential can be calculated.
Further information is available from the manufacturer.
Contact Person
Währinger Straße 38-42
1090
Wien
Room: 3327
Email
Copy email to clipboard
+43-1-4277-73813
Terms of use and usage fees
This device can be used autonomously after successful training.
To be authorised to use this device, you must pass the following tests:
- General Safety Guidelines for Faculty Members
- Laboratory Safety
| User group | Hourly Rate | Cost per sample | Note |
|---|---|---|---|
| Faculties of Physics and Chemistry | - | - | 1. |
| Vienna Life Science Instruments | 2.5 € | 5 € | 2. |
| University users | 5 € | 5 € | 2. |
| Non-university users | 10 € | 5 € |
- For extensive measurement series, the users need to purchase cuvettes.
- Under a publication agreement, the hourly rate is waived and the cost per sample is negotiable.
We also charge for the costs of our staff for training sessions and measurements that cannot be carried out independently.
| User group | Hourly Rate for the Operator | Note |
|---|---|---|
| Faculties of Physics and Chemistry | - | |
| Vienna Life Science Instruments | 70 € | 1. |
| University users | 70 € | 1. |
| Non-university users | 100 € |
- Under a publication agreement, the hourly rate for our employees is also waived.
Specifications
| Device Parameter | Value or Range |
|---|---|
| Measurement Principle | Laser Doppler Electrophoresis |
| Measurement Head | The design of the measuring head used prevents the occurrence of artifacts due to electroosmosis. |
| Zeta-Potential | -500.0 mV bis 500.0 mV |
| Resolution | 0.1 mV (in water) |
| Mobility Range | 10-10 bis 10-7 m2 / V s |
| Temperature Range | 4.0 °C up to 90.0 °C +/- 0.1 °C |
| Maximum Sample Conductivity | 300 mS / m |
| Sample Volumen | 0.5 ml (Standard 4 ml cuvette, glass, PS oder PMMA) |
| Sample Solvents | Compatible with POM |
| Sample Concentration | 0.0001 % bis 10 % (w/w) (depending on the solvent) |
| Particle Size | 1 nm bis 100 µm (not being measured) |
| Laser | 635nm, 20mW |
| Diffration Angle | 17 ° |